欧洲实验室分析用水标准
2016.05.31   点击783次

来源于欧洲EN ISO 3696-1995。

 Parameter Grade1 Grade2 Grade3
 pH value at 25 °C  inclusive range Not applicable Not applicable 5.0 to 7.5
 Electrical conductivity, mS/m at 25℃, max. 0.01 0.1 0.5
 Oxidizable matter Oxygen(O) content mg/l, max Not applicable 0.08 0.4
 Absorbance at 254nm and 1cm optical path
 length, absorbance units, max.
 0.001 0.01 Not specified
 Residue after evaporation on heating at 110℃,
 mg/kg, max.
 Not applicable 1 2
 Silica(SiO2) content mg/l, max. 0.01 0.02 Not specified